Investigating Dose Discrepancies between Raystation TPS and Mobius3D Patient Plan QA Software

Abstract
Purpose

To understand underlying differences in beam modeling leading to occasionally falling gamma analysis criteria

Methods

M3D was commissioned using same beam data as the TPS and default DLG (Dynamic Leaf Gap) for Millenium-120 MLC is used. In about 10% of treatment plans, the gamma index criteria (3mm/3%) falls below 90%. In most of these plans, M3D predicts higher dose in the treated area. These cases have relatively high (> 4) modulation factor (= MU to prescription ratio). To investigate these discrepancies, test plans with (1) small MLC fields, and (2) Alternating open/closed MLC pairs were generated using TPS and compared with M3D.

Results

Good agreement is observed between TPS and M3D along the central axis even for small MLC field sizes (e.g. 1x1cm2 in Fig1). ArcCheck and TPS are in good agreement for open MLC pairs. For all-MLC-closed fields, the dose through rounded MLC end is significantly underestimated by M3D (~60%, Fig.2) but these points represent a small fraction of dose points in gamma analysis, when not masked by jaws. For alternating MLC pairs (Fig.3), we can see that dose corresponding to mid-leaf position (See X dose profile at 1.2cm Y position), the agreement between TPS and M3D is still within a few percent. However, for Y off-axis position corresponding to a leaf edge (e.g. Y = 1.5cm), M3D values become significantly higher (~ 45%).

Conclusion

M3D is a very useful plan QA tool and matches our TPS predictions even for small field sizes. For plans with high modulation factors, an open leaf pair is often neighbored by closed leaves. Therefore, a significant fraction of beamlets delivering dose to the patient are very close to leaf edges and inaccurate modeling of tongue-and-groove effect leads to observed discrepancies. Film dosimetry providing higher spatial resolution is needed for further analysis.

People

Related

Similar sessions

Poster Poster Program
Jul 19 · 07:00
Python-Based Automation Framework for Annual Machine QA Data Archiving In Qatrack+

Annual water-tank measurements help ensure beam characteristics remain consistent with commissioning baselines. However, the lack of a standardized processing workflow and decentralized data storage makes it difficult to analyze...

Syed Bilal Ahmad, PhD
Therapy Physics 0 people interested
Poster Poster Program
Jul 19 · 07:00
User Expectations and Current Availability of HDR Brachytherapy Audits In Europe

The aim of this work was to evaluate the need to implement more dosimetric audits in high‐dose‐rate brachytherapy (HDR-BT) in Europe and to identify which characteristics such audits should meet according to users.

Javier Vijande, PhD Laura Oliver Cañamás
Therapy Physics 0 people interested