Applicator Heterogeneity Effect In Gynecologic HDR Treatments: A Monte Carlo Study

Abstract
Purpose

To physically model the presence of CT/MR-compatible vaginal applicators in HDR dose calculations and accurately quantify the heterogeneity effects on main DVH parameters (D90, D100, V100, and V150).

Methods

HDRMC is an accelerated Monte Carlo dose calculation engine for HDR planning/evaluation. It accounts for tissue and non-tissue heterogeneities. HDRMC was benchmarked against commercial TPS (under the assumption of infinite water medium) and used to re-calculate real patient GYN HDR plans (Chibani and Ma, Med. Phys. 41(5), 2014). In the present study, the dosimetric effects of vaginal applicators with diameters ranging between 20 and 40 mm are investigated. Polyphenylsulfone (PPSU) applicators with 1.29-g/cc density are used. Dose distributions (2-mm voxel grid) are calculated for three models: (1) PPSU applicator surrounded by soft tissue (ICRP46 definition), (2) homogeneous tissue medium, and (3) homogenous water medium. Ten dwell positions are used for each plan

Results

Model 1 (PPSU applicator in tissue) leads to 3% D90 decreases in comparison with Model 3 (homogeneous water) for the 40-mm cylinder. The use of water instead of tissue accounts for about 1% difference. For the 20-mm applicator D90 and D100 differences are smaller (~2%). Accounting for applicator heterogeneity and tissue composition leads to more significant differences in comparison with TPS homogeneous water assumption for V100 (up to 8%) and V150 (up to 28%). See attached data.

Conclusion

The account for applicator heterogeneity and tissue composition for vaginal cylindrical applicators has a rather small effect on D90 and D100 but can lead to significant decrease of V100 and V150.

People

Related

Similar sessions

Poster Poster Program
Jul 19 · 07:00
Python-Based Automation Framework for Annual Machine QA Data Archiving In Qatrack+

Annual water-tank measurements help ensure beam characteristics remain consistent with commissioning baselines. However, the lack of a standardized processing workflow and decentralized data storage makes it difficult to analyze...

Syed Bilal Ahmad, PhD
Therapy Physics 0 people interested
Poster Poster Program
Jul 19 · 07:00
User Expectations and Current Availability of HDR Brachytherapy Audits In Europe

The aim of this work was to evaluate the need to implement more dosimetric audits in high‐dose‐rate brachytherapy (HDR-BT) in Europe and to identify which characteristics such audits should meet according to users.

Javier Vijande, PhD Laura Oliver Cañamás
Therapy Physics 0 people interested