30 Years of End-to-End Dosimetry Audits: The Accuracy of Advanced Radiotherapy

Abstract
Purpose

IROC and Radiation Quality Assurance Lab monitor more than 2500 radiotherapy facilities worldwide to ensure accurate and consistent radiotherapy delivery. This study reviews the 30 year history of the anthropomorphic phantom program for advanced photon therapy.

Methods

The RQALab offers anthropomorphic phantoms for institutions to perform end-to-end tests of the accuracy of their advanced radiotherapy. Phantoms are mailed to the institution where they are treated like a patient, including planning and delivery, and are returned to the RQALab for analysis. Several types of phantoms exist that test IMRT, motion management, SBRT, SRS, multi-target delivery, and more. Institutions may perform this audit as an internal quality assurance test, or may be required to perform this test to participate in clinical trials (e.g., for IROC as part of NCI trials). Passing criteria typically require the dose to be within 5-7% and 3-4mm, and are established in collaboration with the NCI’s clinical trial network.

Results

A total of 9,774 photon phantoms have been irradiated since 1996, most frequently the IMRT H&N phantom (3,429), moving lung phantom (2,788), and SRS phantom (1,898). Overall, the average pass rate is 86.5%, with the highest pass rate seen for the IMRT H&N phantom (89%) and the lowest pass rate seen for the moving multi-target liver phantom (73%). Overall, pass rates have significantly increased with time, which is particularly noteworthy because treatments have gotten more complex over this time period.

Conclusion

This broad evaluation of community practice highlights the progress being made in the face of increasingly complex radiotherapy. However, as the phantom program continues to highlight shortcomings, it also serves as a reminder that constant effort and attention are required to provide the high-level care required by radiotherapy patients.

People

Related

Similar sessions

Poster Poster Program
Jul 19 · 07:00
Python-Based Automation Framework for Annual Machine QA Data Archiving In Qatrack+

Annual water-tank measurements help ensure beam characteristics remain consistent with commissioning baselines. However, the lack of a standardized processing workflow and decentralized data storage makes it difficult to analyze...

Syed Bilal Ahmad, PhD
Therapy Physics 0 people interested
Poster Poster Program
Jul 19 · 07:00
User Expectations and Current Availability of HDR Brachytherapy Audits In Europe

The aim of this work was to evaluate the need to implement more dosimetric audits in high‐dose‐rate brachytherapy (HDR-BT) in Europe and to identify which characteristics such audits should meet according to users.

Javier Vijande, PhD Laura Oliver Cañamás
Therapy Physics 0 people interested