Iroc TBI Phantom Results

Abstract
Purpose

This work reports the development and initial results from irradiations of IROC’s TBI anthropomorphic phantom.

Methods

IROC customized a CIRS 10-year-old phantom by adding 13 TLD locations and building attachable arms from high-impact polystyrene. 3 TLD are located in each lung, one in the brain, one in the neck, one in the mediastinum, one at the umbilicus, one in each kidney, and one in the pelvis. Sites were instructed to deliver 3 Gy to the umbilicus in the manner that they treat patients. If they treated patients with a combination of an open beam and a beam with lung blocks, the phantom was irradiated once with each method. TLD results were compared to the dose predicted by the institution.

Results

The phantom has been irradiated 105 times by 47 institutions since 2017. 12 of the irradiations were VMAT, with the majority of those irradiated since 2023. The remaining were irradiated with opposed fields or sweeping arcs. The phantom was irradiated in a sitting, standing, supine, or decubitus position. Criteria for the traditional non-VMAT irradiations were set as ±5% of the prescription dose at the prescription point and ±15% of the prescription dose for other locations (excluding the lung). Criteria for the VMAT irradiations were set as ±5% of the prescription dose at the prescription point and ±7% of the planned dose for the other non-lung TLD. Due to highly variable doses in the lung, TLD in the lung were not included in the criteria for any treatment. The pass rates were 73% for the traditional irradiations and 92% for the VMAT irradiations. Disagreement exceeded 50% for some points in the lung.

Conclusion

IROC’s TBI phantom shows a variety of traditional TBI techniques as well as increasing use of VMAT TBI irradiations.

People

Related

Similar sessions

Poster Poster Program
Jul 19 · 07:00
Python-Based Automation Framework for Annual Machine QA Data Archiving In Qatrack+

Annual water-tank measurements help ensure beam characteristics remain consistent with commissioning baselines. However, the lack of a standardized processing workflow and decentralized data storage makes it difficult to analyze...

Syed Bilal Ahmad, PhD
Therapy Physics 0 people interested
Poster Poster Program
Jul 19 · 07:00
User Expectations and Current Availability of HDR Brachytherapy Audits In Europe

The aim of this work was to evaluate the need to implement more dosimetric audits in high‐dose‐rate brachytherapy (HDR-BT) in Europe and to identify which characteristics such audits should meet according to users.

Javier Vijande, PhD Laura Oliver Cañamás
Therapy Physics 0 people interested