Poster Poster Program Therapy Physics

Effect of the Spoiler Location on the Total Body Irritation Output

Abstract
Purpose

Total Body Irradiation (TBI) is a crucial component prior to bone marrow transplantation. A beam spoiler is typically used to increase the dose in the high-energy X-ray buildup region to 90% of prescribed dose. This is the first experiment to investigate the effect of spoiler location on TBI dose; it helps determine the optimal spoiler position and improve the consistency of output measurements.

Methods

TBI is implemented using a Varian TrueBeam with the patient in supine position at a nominal treatment distance of 350cm lateral to the target. The treatment parameters include 6MV beam energy, gantry angle of 90°, collimator angle of 0°, 40×40cm field size with MLC retracted. A special block tray is used for mounting compensators. An acrylic beam spoiler measuring 7×5ft with a thickness of 1cm is used. A PTW N30013 Farmer chamber and a Standard Imaging SuperMAX electrometer were used for output measurements. A SunNuclear solid water phantom was placed on a stretcher with the Farmer chamber insert oriented upward so that the chamber axis was perpendicular to the beam from the lateral direction. The chamber measurement depth was 1.5cm.

Results

The spoiler distance was measured relative to TBI nominal treatment distance (350cm) and varied from 50cm to 150cm. The output ratio showed a linear relationship with spoiler distance, increasing from 0.975 at 150cm to 1.01 at 50cm; the closer the spoiler was to the patient, the higher the output.

Conclusion

There are two effects of spoiler: attenuation effect that decreases dose and is dominant at large distances (150cm); and scatter effect that increases the shallow dose and is dominant at short distances (50cm). To achieve consistent dose, 50cm distance was selected, as it is constrained by minimum achievable distance of 48cm, determined by half the lateral width of the stretcher including the turning wheels.

People

Related

Similar sessions

Poster Poster Program
Jul 19 · 07:00
Python-Based Automation Framework for Annual Machine QA Data Archiving In Qatrack+

Annual water-tank measurements help ensure beam characteristics remain consistent with commissioning baselines. However, the lack of a standardized processing workflow and decentralized data storage makes it difficult to analyze...

Syed Bilal Ahmad, PhD
Therapy Physics 0 people interested
Poster Poster Program
Jul 19 · 07:00
User Expectations and Current Availability of HDR Brachytherapy Audits In Europe

The aim of this work was to evaluate the need to implement more dosimetric audits in high‐dose‐rate brachytherapy (HDR-BT) in Europe and to identify which characteristics such audits should meet according to users.

Javier Vijande, PhD Laura Oliver Cañamás
Therapy Physics 0 people interested