The Impact of Accidentally Washing and Drying Your Personal Dosimetry Badges In the Laundry - Does It Matter?

Abstract
Purpose

Occasionally radiation therapy staff accidentally place their personal radiation dosimeter badges in the laundry. This study assessed the effects of standard household laundry wash/dry cycles on the dose readings of Landauer Luxel®+ Optically Stimulated Luminescence (OSL) dosimeter badges.

Methods

Prior to placement in the laundry, dosimeter badges were exposed to scatter and leakage radiation from either a 10 MV or 18 MV photon beam for 0 to 2500 monitor units (MUs) in 500 MU increments, using a clinical TrueBeam linear accelerator. Irradiation conditions were gantry 0o, collimator 0o, field size 20x20 cm2, dose rate 500 MU/min, MLC retracted and nothing in the primary beam. For each exposure, two sets of three badges were placed on a counter, 3 meters away from isocenter, in the treatment room. A radiation survey meter (Victoreen 451P SN1340) and myOSLchip dosimeters (RadPro Int.) were also used to estimate the doses at the measurement location. For each set of exposed badges, one badge was not placed in the laundry, one was placed in a wash cycle (cold water), and one was both washed and dried (medium heat). Two different models of washers and dryers were used for the two badge sets. After washing and drying, all badges were sent to Landauer® for reading. Survey meter readings were used to correct for slight variations in dose between the badge locations on the counter.

Results

Badge readings differed from myOSLchip and survey meter readings by up to 0.12 mSv and 0.06 mSv respectively. Washing and drying the badges had no noticeable effect on the dose readings (maximum difference 0.07 mSv). There was no clear trend observed across test conditions.

Conclusion

Luxel®+ OSL dosimeter badge readings were unaffected by laundry washing and drying. If not physically damaged, replacement of accidentally washed and dried badges may not be necessary.

People

Related

Similar sessions

Poster Poster Program
Jul 19 · 07:00
Python-Based Automation Framework for Annual Machine QA Data Archiving In Qatrack+

Annual water-tank measurements help ensure beam characteristics remain consistent with commissioning baselines. However, the lack of a standardized processing workflow and decentralized data storage makes it difficult to analyze...

Syed Bilal Ahmad, PhD
Therapy Physics 0 people interested
Poster Poster Program
Jul 19 · 07:00
User Expectations and Current Availability of HDR Brachytherapy Audits In Europe

The aim of this work was to evaluate the need to implement more dosimetric audits in high‐dose‐rate brachytherapy (HDR-BT) in Europe and to identify which characteristics such audits should meet according to users.

Javier Vijande, PhD Laura Oliver Cañamás
Therapy Physics 0 people interested